首页> 外文期刊>The Journal of hand surgery, European volume >A knotless bidirectional-barbed tendon repair is inferior to conventional 4-strand repairs in cyclic loading
【24h】

A knotless bidirectional-barbed tendon repair is inferior to conventional 4-strand repairs in cyclic loading

机译:A knotless bidirectional-barbed tendon repair is inferior to conventional 4-strand repairs in cyclic loading

获取原文
获取原文并翻译 | 示例
           

摘要

We divided 21 flexor digitorum profundus tendons in the index, middle and ring fingers in seven cadaver hands into three groups. The tendons were cut in zone 2 and repaired using a 4-strand cruciate core suture repair with one of the following three materials in each group: (1) a knotless repair with a 2-0 bidirectional-barbed suture, which has similar tensile strength as a 4-0 non-barbed suture used in the other two groups; (2) a knotted locking repair with a non-barbed 4-0 conventional suture; and (3) a non-locking repair with a non-barbed 4-0 knotless suture. The repaired fingers were cyclically loaded through a simulated active range of motion to a 5N load. We monitored and recorded the gap sizes at regular intervals during the test. The 2-0 bidirectional-barbed suture group and non-barbed suture groups developed gaps of 2.2mm after 10 cycles and 2.4mm after 20 cycles, respectively. Over 1000 cycles, the mean gaps were 3.2mm in the 4-0 conventional suture group and 9.1mm in the 2-0 bidirectional-barbed group. The tendons in the 2-0 bidirectional-barbed group gapped earlier, with statistically significant differences compared with those in the locking repair with a non-barbed 4-0 knotless suture group. The repair strength of the barbed suture technique was inferior to the cruciate repairs using a conventional 4-0 non-barbed suture tested in this cyclic-loading model. Level of evidence: Level V

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号