首页> 外文期刊>investigative radiology >Computed Tomographic Volumetric Calculation Reproducibility
【24h】

Computed Tomographic Volumetric Calculation Reproducibility

机译:计算机断层扫描体积计算重现性

获取原文
获取外文期刊封面目录资料

摘要

Previous studies have described computed tomographic (CT) measurement of the total volume of large organs and tumors. This study examines sources of error in the CT volumetric measurement of small anatomic structures, and has potential application to the volumetric measurement of small extravisceral tumors.Volume is determined by multiplying observer measurement of area by the CT slice thickness. We focused on the measurement of small cross-sectional areas as a major source of error in volumetric calculations. One observer made ten area measurements on each of 12 structures of various sizes. For areas larger than 8 to 10 cm2, any single measurement was within 2 to 4 of the mean of ten values. For areas smaller than 8 to 10 cm2, measurements varied more with respect to the mean, and rose approximately exponentially as the measured cross-sectional area approached zero.
机译:以前的研究已经描述了大器官和肿瘤总体积的计算机断层扫描 (CT) 测量。本研究探讨了小解剖结构的CT体积测量的误差来源,并具有潜在的应用,可用于小的内脏外肿瘤的体积测量。体积是通过观察者测量面积乘以 CT 切片厚度来确定的。我们专注于小横截面积的测量,这是体积计算中误差的主要来源。一名观察者对 12 个不同尺寸的结构中的每一个进行了 10 次面积测量。对于大于 8 至 10 cm2 的区域,任何单个测量值都在 10 个值的平均值的 2 到 4% 以内。对于小于 8 至 10 cm2 的区域,测量值相对于平均值的变化更大,并且随着测量的横截面积接近零,测量值呈指数级上升。