Significant RF-EMF and thermal levels observed in a computational model of a person with a tibial plate for grounded 40 MHz exposure

机译:在胫骨板接地 40 MHz 暴露的人的计算模型中观察到显着的 RF-EMF 和热水平

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Using numerical modeling, a worst-case scenario is considered when a person with a metallic implant is exposed to a radiofrequency (RF) electromagnetic field (EMF). An adult male standing on a conductive ground plane was exposed to a 40MHz vertically polarized plane wave field, close to whole-body resonance where maximal induced current flows are expected in the legs. A metal plate (50-300mm long) was attached to the tibia in the left leg. The findings from this study re-emphasize the need to ensure compliance with limb current reference levels for exposures near whole-body resonance, and not just rely on compliance with ambient electric (E) and magnetic (H) field reference levels. Moreover, we emphasize this recommendation for someone with a tibial plate, as failure to comply may result in significant tissue damage (increases in the localized temperature of 5-10°C were suggested by the modeling for an incident E-field of 61.4V/m root mean square (rms)). It was determined that the occupational reference level for limb current (100mArms), as stipulated in the 1998 guidelines of the International Commission on Non-Ionizing Radiation Protection (ICNIRP), is satisfied if the plane wave incident E-field levels are no more than 29.8V/mrms without an implant and 23.4V/mrms for the model with a 300mm implant. Bioelectromagnetics. 35:284-295, 2014.
机译:使用数值建模,当金属植入物的人暴露于射频 (RF) 电磁场 (EMF) 时,考虑了最坏的情况。一名成年男性站在导电接地平面上,暴露在40MHz的垂直极化平面波场中,接近全身共振,预计腿部会有最大的感应电流。一块金属板(50-300毫米长)附着在左腿的胫骨上。这项研究的结果再次强调,需要确保在接近全身共振的暴露中符合肢体电流参考水平,而不仅仅是依赖于对环境电 (E) 和磁场 (H) 参考水平的遵守。此外,我们强调对胫骨板患者的建议,因为不遵守可能会导致严重的组织损伤(入射均方根 (rms) 的入射电场建模表明局部温度升高 5-10°C)。确定,如果平面波入射电场水平不超过 29.8V/mrms(没有植入物)和 23.4V/mrms(带 300mm 植入物的型号)不超过 23.4V/mrms,则满足国际非电离辐射防护委员会 (ICNIRP) 1998 年指南中规定的肢体电流 (100mArms) 的职业参考水平。生物电磁学。35:284-295, 2014.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号