Comparison of Fluorescence Polarization Immunoassay with Liquid Chromatography for Quantification of Procainamide andN-Acetylprocainamide in Urine

机译:Comparison of Fluorescence Polarization Immunoassay with Liquid Chromatography for Quantification of Procainamide andN-Acetylprocainamide in Urine

获取原文
           

摘要

SummaryThe objective of this study was to compare the precision and accuracy of fluorescence polarization immunoassay (FPIA) with high-performance liquid chromatography (HPLC) for measurement of procainamide (PA) andN-acetylprocainamide (NAPA) concentrations in urine. To determine the correlation between FPIA and HPLC, urine PA and NAPA concentrations were assayed using both techniques in samples obtained from study patients receiving PA and in spiked samples. In samples from patients, FPIA-determined PA and NAPA concentrations were 19 plusmn; 9 lower and 28 plusmn; 31 higher, respectively, than those determined by HPLC. The slope of the FPIA-HPLC regression lines for PA and NAPA differed significantly from that of the line of unity (the slope that would result if FPIA and HPLC yielded identical concentrations). In spiked samples, FPIA-determined PA and NAPA concentrations were 15 plusmn; 2 and 11 plusmn; 2 lower than HPLC-determined concentrations, respectively, and the slopes of the FPIA-HPLC regression lines differed significantly from the line of unity. Therefore, FPIA cannot be recommended as a urine assay method when quantitative assessment of urine PA or NAPA excretion is needed for pharmacokinetic studies.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号