Metallosis caused by alumina ceramic screw fixation of grafted bone in a bipolar hip endoprosthesis

机译:双极髋关节内假体中移植骨的氧化铝陶瓷螺钉固定引起的金属病

获取原文
获取外文期刊封面目录资料

摘要

We report a case of metallosis due to ceramic fragments in the inner bearing surface of the polyethylene cup of a bipolar hip prosthesis with fracture of the alumina ceramic screws used for grafted bone fixation. After bone-grafting acetabuloplasty with a Bateman bipolar endoprosthesis, gradual migration of the outer head occurred, and fracture of the screws was observed. Seven years after the operation, radiographs showed severe metallosis of the surrounding tissues. At revision, marked wear of the inner head was seen. Detailed examination revealed that many rough-edged ceramic particles originating from the fractured ceramic screws had become embedded in the inner surface of the polyethylene bearing insert. These extremely hard ceramic fragments had entered the prosthetic joint space during joint movement, thereafter causing continuous abrasion of the metal head. After revision, the concentration of the constituents of the prosthesis alloy in the serum decreased toward control levels. To prevent serious complications like those in the present case, screws should be inserted more than 1 cm from the outer head of a bipolar endoprosthesis, and the outer head should be removed as soon as possible if it comes into contact with the screws.
机译:我们报告了一例由于双极髋关节假体聚乙烯杯内承载表面的陶瓷碎片而导致的金属病病例,用于移植骨固定的氧化铝陶瓷螺钉断裂。用贝特曼双极内假体进行骨移植髋臼成形术后,外头逐渐迁移,并观察到螺钉骨折。手术七年后,X光片显示周围组织严重金属化。在修订时,可以看到内头部的明显磨损。详细检查显示,许多来自断裂陶瓷螺钉的粗糙陶瓷颗粒已嵌入聚乙烯轴承嵌件的内表面。这些极其坚硬的陶瓷碎片在关节运动过程中进入了假体关节空间,随后导致金属头不断磨损。修订后,血清中假体合金成分的浓度向对照水平下降。为防止出现像本案这样的严重并发症,螺钉应插入距离双极内假体外头 1 厘米以上的地方,如果外头与螺钉接触,应尽快取下。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号