首页> 外文期刊>Archives of orthopaedic and trauma surgery. >Fixation of a modular curved revision stem with a taper of 2C in the femur
【24h】

Fixation of a modular curved revision stem with a taper of 2C in the femur

机译:在股骨中固定锥度为 2C 的模块化弯曲翻修杆

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Introduction Modular revision stems with a short distal component can prevent the bypassing of the femoral isthmus and hereby theoretically have advantages concerning risk of periprosthetic fractures, breakage of the junction and a technically easier revision procedure. Materials and methods Radiological evaluation of 202 stem revision operations with the modular curved revision stem "Revi- tan Curved" with a 2° taper was carried out after a mean follow-up period of 7.44 ± 2.09 years (3-13 years) to investigate whether short-stem combinations are effective in Paprosky 2 and 3A defects with respect to rate of subsidence and loosening. Results Sixty of 62 endofemoral (96.8) and 137 of 140 transfemoral implantations (97.9) involved the short, 140 mm distal component. Significant subsidence was seen in 3.3 of cases following endofemoral implantation and in 2.1 of cases following transfemoral implantation. Neither aseptic loosening nor periprosthetic fracture were observed. Conclusion The use of combinations of short modular components leads to reproducibly good outcomes in femoral revision with respect to subsidence and loosening.
机译:引言 具有短远端组件的模块化翻修杆可以防止股骨峡部的旁路,因此理论上具有假体周围骨折、连接处断裂风险和技术上更容易的翻修程序的优势。材料和方法 在平均随访期为 7.44 ± 2.09 年(3-13 年)后,对 202 次使用具有 2° 锥度的模块化弯曲翻修茎“Revi- tan Curved”的阀杆翻修操作进行了放射学评估,以研究短阀杆组合在沉降和松动速率方面对 Paprosky 2 和 3A 缺陷是否有效。结果 62 例股内植入物中有 60 例 (96.8%) 和 140 例经股骨植入物中有 137 例 (97.9%) 涉及 140 mm 的短远端部分。3.3% 的股内植入后病例出现显著沉降,2.1% 的经股骨植入后出现显著沉降。未观察到无菌性松动和假体周围骨折。结论 使用短模块化组件的组合在股骨修复中可重复地改善下沉和松动。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号