首页> 外文期刊>Analytical and bioanalytical chemistry >Determination of albumin in biological fluids by flow injection analysis using the peroxyoxalate chemiluminescent system in micellar medium
【24h】

Determination of albumin in biological fluids by flow injection analysis using the peroxyoxalate chemiluminescent system in micellar medium

机译:Determination of albumin in biological fluids by flow injection analysis using the peroxyoxalate chemiluminescent system in micellar medium

获取原文
获取原文并翻译 | 示例
           

摘要

A new sensitive analytical procedure for the determination of albumin is presented, based on the participation of a fluorescent derivative of the protein in the peroxyoxalate chemiluminescent system using imidazole as a catalyst. The chemiluminescent emission is detected in a flow injection analysis (FIA) system, employing micellar medium as carrier, whereas the derivatisation reaction is carried out off-line using fluorescamine as a label. The optimisation of the operational and chemical variables of the method such as concentration of reagents, flow-rates, injection volume, etc., involves the use of experimental designs, including Draper-Lin small composite design, scarcely used in analytical chemistry. The performance characteristics were established and a detection limit of about 0.1 fmol of albumin was obtained. The method has been validated and satisfactorily applied to the determination of albumin in human serum and urine.