Accuracy and Reliability of Uterine Contraction Identification Using Abdominal Surface Electrodes

机译:Accuracy and Reliability of Uterine Contraction Identification Using Abdominal Surface Electrodes

获取原文
获取原文并翻译 | 示例
           

摘要

Objective: To compare the accuracy and reliability of uterine contraction identification from maternal abdominal electrohysterogram and tocodynamometer with an intrauterine pressure transducer.Methods: Seventy-four term parturients had uterine contractions monitored simultaneously with electrohysterography, tocodynamometry, and intrauterine pressure measurement.Results: Electrohysterography was more reliable than tocodynamometry when compared to the intrauterine method (97.1 versus 60.9 positive percent agreement; P < 0.001). The root mean square error was lower for electrohysterography than tocodynamometry in the first stage (0.88 versus 1.22 contractions/10 minutes; P < 0.001), and equivalent to tocodynamometry in the second. The positive predictive values for tocodynamometry and electrohysterography (84.1 versus 78.7) were not significantly different, nor were the false positive rates (21.3 versus 15.9; P = 0.052). The sensitivity of electrohysterography was superior to that of tocodynamometry (86.0 versus 73.6; P < 0.001).Conclusion: The electrohysterographic technique was more reliable and similar in accuracy to tocodynamometry in detecting intrapartum uterine contractions.