首页> 外文期刊>international journal of imaging systems and technology >Analysis of MTF degradation in the imaging of cells in a flow system
【24h】

Analysis of MTF degradation in the imaging of cells in a flow system

机译:Analysis of MTF degradation in the imaging of cells in a flow system

获取原文
           

摘要

AbstractThe time‐delay and integration method can be used with a charge‐coupled device array to acquire images of cell in a flow cytometer. The loss of high spatial frequency components in images acquired with this technique is studied by deriving the modulation transfer function (MTF) for each of the main causes of image degradation: finite sampling aperture, nonsynchronism between image movement and charge transfer rate, discrete charge motion, transfer inefficiency, and axis misalignment. The system MTF and resolution in the horizontal and vertical directions are then examined.©1994 John WileySon