首页> 外文期刊>annals of plastic surgery >Complications and Removal Rates of Miniplates and Screws Used for Maxillofacial Fractures
【24h】

Complications and Removal Rates of Miniplates and Screws Used for Maxillofacial Fractures

机译:Complications and Removal Rates of Miniplates and Screws Used for Maxillofacial Fractures

获取原文
           

摘要

Complications of miniplates and screws used for maxillofacial fractures were analyzed, and complications were evaluated in relation to fracture site. Motor vehicle accidents were the cause of all fractures in this study. During the last 7 years (1994ndash;2001), noncompressive titanium miniplates and screws were used for stabilization of maxillofacial fractures. In 66 patients, 87 fracture sites were stabilized using 296 miniplates and 1,184 screws. The mean age of the patients was 31 years (age range, 6ndash;64 years). The percentage of male patients was 77percnt; and the percentage of female patients was 23percnt;. Miniplates and screws were used in 6 patients (10percnt;) who were younger than 15 years of age at the time of the surgery. The follow-up period ranged between 3 months and 7 years. The overall miniplate and screw removal rate was 7percnt;. The rates of removal according to the fracture site are as followscolon; mandible, 4.4percnt;; zygomaticofrontal junction, 1.4percnt;; inferior orbital rim, 0.7percnt;; maxilla, 0.3percnt;; and frontal sinus wall, 0.3percnt;. Removal causes were infection, 2percnt;; extrusion, 1.7percnt;; visibility, 1.4percnt;; pain, 1percnt;; malunion, 0.7percnt;; and miniplate fracture, 0.3percnt;. The minimum time period between insertion and removal was 3 months and the maximum period was 14 months. Infection and extrusion were the main complications for removal of miniplates and screws from the mandible, whereas miniplates and screws were removed from the zygoma because of visibility (zygomaticofrontal region) under the skin in the vast majority of the patients. The maxilla was the least operated region for miniplate and screw removal. In all patients in this study, the preoperative physical symptoms were relieved after miniplate and screw removal. Miniplates and screws are very useful tools in maxillofacial fracture management, but sometimes they have to be removed. In the authorsrsquo; series, the removal rate was 7percnt;, and this rate can vary with the severity of the trauma and location of the fracture.Islamoglu K, Coskunfirat OK, Tetik G, Ozgentas HE. Complications and removal rates of miniplates and screws used for maxillofacial fractures.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号