AbstractA rapid method for the determination of 4‐methylpyrazole in plasma is described. Internal standard (3‐methylpyrazole) is added to the plasma which is subsequently applied to a BondElut SCX column. After washing the column the 3‐ and 4‐methylpyrazoles are eluted with a phosphate buffer and analyzed by isocratic reversed‐phase high performance liquid chromatography with UV detection. The concentration of 4‐methylpyrazole is calculated from the 4‐methylpyrazole/3‐methylpyrazole peak high ratio. The method is linear between 2.5 and 100 μmol/L with a within‐day precision (CV) of 2.2 (n= 10) and a day‐to‐day precision of 2.8 (n= 30). The sensitivity is sufficient for analysis of plasma levels in th