首页> 外文期刊>spine >Biomechanical Evaluation of Spinal Fixation DevicesPart III. Stability Provided by Six Spinal Fixation Devices and Interbody Bone Graft
【24h】

Biomechanical Evaluation of Spinal Fixation DevicesPart III. Stability Provided by Six Spinal Fixation Devices and Interbody Bone Graft

机译:Biomechanical Evaluation of Spinal Fixation DevicesPart III. Stability Provided by Six Spinal Fixation Devices and Interbody Bone Graft

获取原文
           

摘要

The three-dimentional stability provided by six spinal fixation devices with or without interbody bone graft has been studied in an in vitro biomechanical model using five-vertebral (T11-L3) fresh cadaveric thoracolumbar specimens. An injury was created at T12-L1 by complete transection of the posterior elements and posterior half of the intervertebral disc, leaving the anterior half of the intervertebral disc and anterior longitudinal ligament intact. The three-dimensional rotations and translations, measures of biomechanical instabilities, were determined under physiologic loads for the intact spine and the spinal constructs, ie, injured spine plus instrumentation. The tested devices werecolon; Harrington reverse ratchet rods (HR); Luque rectangle rod (LR); Kaneda device without transverse fixator (KD); Kaneda device with transverse fixators (KT); transpedicular external fixator (EF). In addition, stability tests were performed for KT, EF, and Harrington compression rods with interbody bone graft following a corpectomy (KTB, EFB, and HCB). The constructs were more stable than the intact spine under the four loads in the following ordercolon; flexioncolon; EFB, HCB, EF, HR, LR, KTB, and KT; extensioncolon; EFB, LR, EF, KTB, HR, and KT; lateral bendingcolon; KTB, KT, EFB, KD, EF, HCB, and HR; and axial rotationcolon; EFB.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号