PET/CT imaging: Effect of respiratory motion on apparent myocardial uptake

机译:PET/CT imaging: Effect of respiratory motion on apparent myocardial uptake

获取原文
获取原文并翻译 | 示例
           

摘要

Background. Positron emission tomography (PET) attenuation correction (AC) using computed tomography (CT) can be affected by respiratory motion: hi-speed CT captures 1 point of the respiratory cycle while PET emission data averages many cycles. We quantified the changes hi apparent myocardial uptake due to this respiratory-induced CT attenuation mismatch.Methods. Twenty-two patients undergoing fluorme-18 fluorodexyglucose (FDG) PET/CT received 3 sequential CT scans at normal resting end-inspiration (CTINSPIR), ending expiration (CTEXPIR), and at midvolume between end-expiration and end-inspiration (CT). A pneumotachometer measured absolute changes in lung volume. Seven subjects also underwent a 3-minute transmission scan with a 68Ge rotating rod source (RRS). The PET emission data set was reconstructed up to 4 tunes using CTEXPIR, CTINSP1R, CTMIDVOL, and RRS AC maps. Relative heart position and cardiac uptake was measured for each CT attenuation correction.Results. Respiratory motion produced marked changes in global and regional myocardial uptake; Changes were large hi the lateral and anterior regions at the lung-soft tissue interface (up to 30 using CTmsP1R compared to CTEXPIK for AC) and smaller in the septal region (10 or less). Data corrected with CTEXPIR agreed best with the RRS.Conclusion. Respiratory effects can introduce large inhomogeneities hi apparent myocardial uptake when CT is used for attenuation correction. (J Nucl Cardiol 2006;13:821-30.)

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号