首页> 外文期刊>Optoelectronics letters >Essential parameter calibration for the 3D scanner with only single camera and projector
【24h】

Essential parameter calibration for the 3D scanner with only single camera and projector

机译:Essential parameter calibration for the 3D scanner with only single camera and projector

获取原文
获取原文并翻译 | 示例
           

摘要

A calibration method for the five essential parameters is proposed. Using the calibration results, the three dimensional (3D) reconstruction can be performed directly. The five essential parameters include the distance between the camera and the projector, the distance between the reference plane and the camera, the fundamental frequency of the fringe pattern, the scale factor from the image coordinates to the world coordinate system in X axis direction and that in Y axis direction. The proposed calibration method is implemented and tested in our 3D reconstruction system. The mean calibration error is found to be 0. 0215 mm over a volume of 400 mm (H)×300 mm (V)×500 mm (D). The proposed calibration method is accurate and useful for the 3D reconstruction system.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信