首页> 外文期刊>Analytical and bioanalytical chemistry >Liquid chromatography-tandem mass spectrometry analysis of 17-hydroxyprogesterone in dried blood spots revealed matrix effect on immunoassay
【24h】

Liquid chromatography-tandem mass spectrometry analysis of 17-hydroxyprogesterone in dried blood spots revealed matrix effect on immunoassay

机译:Liquid chromatography-tandem mass spectrometry analysis of 17-hydroxyprogesterone in dried blood spots revealed matrix effect on immunoassay

获取原文
获取原文并翻译 | 示例
           

摘要

Immunoassays for measuring 17-hydroxyprogesterone (17-OHP) produce high rates of false positives that impact the identification of congenital adrenal hyperplasia (CAH) in neonates. A confirmatory test with high analytical specificity employing liquid chromatography-tandem mass spectrometry (LC-MS/MS) methodology is needed in newborn screening for CAH. 17-OHP and cortisol were extracted from dried blood spot (DBS) samples, resolved on a C18 column, and measured using tandem mass spectrometry. The results were compared with those determined using the AutoDELFIA immunoassay. The LC-MS/MS method had a limit of quantitation of 10.0 and 5.0ng/mL for 17-OHP and cortisol, respectively. The method characteristics showed coefficient variation (CV) 11.9 for both 17-OHP and cortisol, recoveries ranging from 83.1 to 101.5 for 17-OHP and from 95.1 to 102.8 for cortisol, and linearity with R-2=0.9994 for 17-OHP and R-2=0.9996 for cortisol, clinical sensitivity of 100.0 and a specificity of 96.4 as obtained by receiver operating characteristic analysis on 45 patient samples when 17-OHP 39.1ng/mL was selected as the cutoff value. Comparison between the LC-MS/MS and the AutoDELFIA immunoassay methods revealed a poor correlation for patient DBS samples (R-2=0.6784); however, an excellent correlation was obtained for QC and proficiency test (PT) DBS samples (R-2=0.9797). The LC-MS/MS method produced reliable results for 17-OHP and cortisol for the diagnosis of CAH. The AutoDELFIA immunoassay appears to be subject to matrix effects in the analysis for 17-OHP in DBS patient samples. The DBS samples of non-patient origin may not be suitable for assessing analytical accuracy of immunoassays.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号