首页> 外文期刊>The Journal of hand surgery, European volume >High revision rates with the metal-on-metal Motec carpometacarpal joint prosthesis
【24h】

High revision rates with the metal-on-metal Motec carpometacarpal joint prosthesis

机译:High revision rates with the metal-on-metal Motec carpometacarpal joint prosthesis

获取原文
获取原文并翻译 | 示例
           

摘要

We retrospectively evaluated a consecutive series of 42 Motec thumb carpometacarpal total joint arthroplasties. The primary endpoint was revision with implant removal and trapeziectomy. At follow-up the disability of the arm shoulder and hand (DASH) score, pain on numerical rating scale at rest and with activity and serum chrome and cobalt concentrations were assessed for both unrevised and revised patients. At a mean follow-up of 26 months, 17 patients had been revised. The 2 year cumulative revision rate was 42 (95 CI, 28-60). The DASH score and pain scores at rest and with activity were comparable between the patients whose thumbs remained unrevised and those revised. Patients with elevated serum chrome and cobalt levels had significantly higher DASH and pain scores, but elevated levels were not associated with revision. The revision rate in this study is unacceptably high. However, pain and DASH scores after revision are acceptable and comparable with patients with non-revised implants.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有