首页> 外文期刊>artificial organs >The Dura Mater Valve:In VitroCharacteristics and Pathological Changes After Implantation in Calves
【24h】

The Dura Mater Valve:In VitroCharacteristics and Pathological Changes After Implantation in Calves

机译:硬脑膜:体外特征及犊牛植入后的病理变化

获取原文
获取外文期刊封面目录资料

摘要

ABSTRACTHuman dura mater valves of various sizes with rigid and flexible stents were tested in anin vitropulsatile mock circulatory system. A 22‐mm flexible stent valve incorporating a new fabrication technique showed almost the same pressure gradient as a 28‐mm rigid stent valve. The backflow/stroke volume ratio was about 4 at a net flow of 10 L/min. One hundred and five rigid stent‐mounted dura mater valves were used in 51 pump implantations for up to 316 days. Collagen fiber degeneration began three months after implantation. Microscopic and macroscopic calcification of the valve tissue was seen in eight out of 105 valves, giving an overall incidence of 7.6. The calcified degeneration was dystrophic in nature, not accompanied by cellular reactions, and was seen in the areas of the valve under stress. The degenerative changes were more severe in the left side than in the right side of the total artificial heart. These findings suggest that mechanical damage to the tissue plays an important role in the pathogenesis of calcific
机译:摘要在体外轻动模拟循环系统中测试了各种尺寸的硬脑膜和柔性支架。采用新制造技术的 22 毫米柔性支架阀显示出与 28 毫米刚性支架阀几乎相同的压力梯度。在10 L/min的净流量下,回流/冲程容积比约为4%。105 个刚性支架安装硬脑膜瓣膜用于 51 次泵植入,持续时间长达 316 天。胶原纤维变性在植入后三个月开始。在 105 个瓣膜中,有 8 个瓣膜组织出现微观和宏观钙化,总发生率为 7.6%。钙化变性本质上是营养不良的,不伴有细胞反应,可见于应力下的瓣膜区域。全人工心脏左侧的退行性变化比右侧更严重。这些发现表明,组织的机械损伤在钙化的发病机制中起着重要作用