Multicenter study of oxygen-insensitive handheld glucose point-of-care testing in critical care/hospital/ambulatory patients in the United States and Canada

机译:Multicenter study of oxygen-insensitive handheld glucose point-of-care testing in critical care/hospital/ambulatory patients in the United States and Canada

获取原文
           

摘要

ObjectivesExisting handheld glucose meters are glucose oxidase (GO)-based. Oxygen side reactions can introduce oxygen dependency, increase potential error, and limit clinical use. Our primary objectives were to: a) introduce a new glucose dehydrogenase (GD)-based electrochemical biosensor for point-of-care testing; b) determine the oxygen-sensitivity of GO- and GD-based electrochemical biosensor test strips; and c) evaluate the clinical performance of the new GD-based glucose meter system in critical care/hospital/ambulatory patients.DesignMulticenter study sites compared glucose levels determined with GD-based biosensors to glucose levels determined in whole blood with a perchloric acid deproteinization hexokinase reference method. One site also studied GO-based biosensors and venous plasma glucose measured with a chemistry analyzer. Biosensor test strips were used with a handheld glucose monitoring system. Bench and clinical oxygen sensitivity, hematocrit effect, and precision were evaluated.SettingThe study was performed at eight U.S. medical centers and one Canadian medical center.PatientsThere were 1,248 patients.Results100 mg/dL, compared with the whole-blood reference method results. With the GD-based biosensor, the percentages of glucose measurements that were not within the error tolerance were comparable for different specimen types and clinical groups. Bracket predictive values were acceptable for glucose levels used in therapeutic management.ConclusionsThe performance of GD-based, oxygen-insensitive, handheld glucose testing was technically suitable for arterial specimens in critical care patients, cord blood and heelstick specimens in neonates, and capillary and venous specimens in other patients. Multicenter findings benchmark the performance of bedside glucose testing devices. With the new +/- 15 mg/dL right arrow 100 mg/dL right arrow +/- 15 accuracy criterion, point-of-care systems for handheld glucose testing should score 95 (or better), as compared with the recommended reference method. Physiologic changes, preanalytical factors, confounding variables, and treatment goals must be taken into consideration when interpreting glucose results, especially in critically ill patients, for whom arterial blood glucose measurements will reflect systemic glucose levels. (Crit Care Med 1998; 26:581-590)

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号