首页> 外文期刊>artificial organs >Comparison of Chemical Analyses of Hollow‐Fiber Dialyzer Extracts
【24h】

Comparison of Chemical Analyses of Hollow‐Fiber Dialyzer Extracts

机译:中空纤维透析器提取物的化学分析比较

获取原文

摘要

Abstract:Hollow‐fiber hemodialyzers containing cellulose‐based membranes have been shown to produce positive results with theLimulusamebocyte lysate (LAL) test. This study was undertaken to determine whether endotoxin was causing the reaction. Rinses from 45 parallel‐plate and hollow‐fiber dialyzers from eight different manufacturers were tested using three lysates and four LAL methods. In addition, four in vitro cellular methods‐human leukocytic pyrogen, lymphocytic activating factor, peritoneal macrophage, and arginase release‐were used to evaluate endotoxin activity. The substance causing the reaction was identified using chromatographic methods. Results indicate that the LAL‐reactive material is cellulose derived and is n
机译:摘要:含有纤维素基膜的中空纤维血液透析器在Limulusamebocyte lysate (LAL)测试中显示出阳性结果。进行这项研究是为了确定内毒素是否引起反应。来自 8 家不同制造商的 45 台平行板和中空纤维透析器的冲洗液使用三种裂解物和四种鲎剂方法进行了测试。此外,四种体外细胞方法——人白细胞热原、淋巴细胞活化因子、腹膜巨噬细胞和精氨酸酶释放——用于评估内毒素活性。使用色谱法鉴定引起反应的物质。结果表明,鲎试剂反应性材料是纤维素衍生的,并且是n