High Pressure Liquid Chromatography for Simultaneous Analysis of AnticonvulsantsComparison with EMITreg;System

机译:High Pressure Liquid Chromatography for Simultaneous Analysis of AnticonvulsantsComparison with EMITreg;System

获取原文
       

摘要

Summarycolon;We present a new high pressure liquid chromatography lpar;HPLCrpar; method for the simultaneous analysis of primidone, phenobarbital, phenytoin, and carbamazepine in serum. The chromatographic separation is carried out using an Altex model 110hyphen;A pump, a 250 times; 4.6 mm column containing 5 mgr;m Spherisorb ODS particles and a variable wavelength ultraviolet detector set at 197 nm. The mobile phase is a mixture of acetonitrile, distilled water, and 1.75 M phosphoric acid lpar;27colon;72.8colon;0.2rpar;. The flow rate is 1.5 mlsol;min, and the analysis time is 17 min. A 200 mgr;l aliquot of serum is buffered at pH 5 and extracted with dichloromethane. The extract is evaporated to dryness and dissolved in methanol for chromatographic analysis. Cyclopal is used as the internal standard and quantification is achieved using peak height ratios. This HPLC method is evaluated for precision and accuracy with reference to the EMITreg;system. The leasthyphen;squares regression analysis of comparison data for the drugs shows a favorable correlation. Also, a pairedthyphen;test indicates no significant difference for the HPLC and EMIT values for primidone, phenobarbital, phenytoin, and carbamazepine. From this study we conclude that this HPLC method could be successfully used for the simultaneous therapeutic monitoring of the four anticonvulsants.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号