Dosimetric impact of a robust optimization approach to mitigate effects from rotational uncertainty in prostate intensity‐modulated brachytherapy

机译:Dosimetric impact of a robust optimization approach to mitigate effects from rotational uncertainty in prostate intensity‐modulated brachytherapy

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Abstract Background Intensity‐modulated brachytherapy (IMBT) is an emerging technology for cancer treatment, in which radiation sources are shielded to shape the dose distribution. The rotatable shields provide an additional degree of freedom, but also introduce an additional, directional, type of uncertainty, compared to conventional high‐dose‐rate brachytherapy (HDR BT). Purpose We propose and evaluate a robust optimization approach to mitigate the effects of rotational uncertainty in the shields with respect to planning?criteria. Methods A previously suggested prototype for platinum‐shielded prostate 169Yb‐based dynamic IMBT is considered. We study a retrospective patient data set (anatomical contours and catheter placement) from two clinics, consisting of six patients that had previously undergone conventional 192Ir HDR BT treatment. The Monte Carlo‐based treatment planning software RapidBrachyMCTPS is used for dose calculations. In our computational experiments, we investigate systematic rotational shield errors of ±10° and ±20°, and the same systematic error is applied to all dwell positions in each scenario. This gives us three scenarios, one nominal and two with errors. The robust optimization approach finds a compromise between the average and worst‐case scenario?outcomes. Results We compare dose plans obtained from standard models and their robust counterparts. With dwell times obtained from a linear penalty model (LPM), for 10° errors, the dose to urethra (D0.1cc$D_{0.1cc}$) and rectum (D0.1cc$D_{0.1cc}$ and D1cc$D_{1cc}$) increase with up to 5 and 7, respectively, in the worst‐case scenario, while with the robust counterpart, the corresponding increases were 3 and 3. For all patients and all evaluated criteria, the worst‐case scenario outcome with the robust approach had lower deviation compared to the standard model, without compromising target coverage. We also evaluated shield errors up to 20° and while the deviations increased to a large extent with the standard models, the robust models were capable of handling even such large?errors. Conclusions We conclude that robust optimization can be used to mitigate the effects from rotational uncertainty and to ensure the treatment plan quality of?IMBT.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号