The Objective Measurement and Subjective Perception of Flexible ENT Endoscopes' Image Quality

机译:柔性耳鼻喉内窥镜图像质量的客观测量与主观感知

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

ENT-flexible endoscopes are an important tool for ear, nose and throat (ENT) professionals to examine the upper airway. Although image quality has improved significantly in the past decade, there is no generally accepted approach to measure this objectively. Sharpness, visual noise and color fidelity are aspects of image quality that can objectively be measured. The purpose of this study was to explore the relationship between these quality metrics and the subjective perception of image quality by ENT-professionals. The image quality of six different flexible endoscopes was assessed objectively and subjectively. Objective measurements were obtained using the Rez Checker Target Nano Matte and comprised sharpness (MTF50), visual noise and color fidelity (CIE Delta E 2000). Subjective image quality ranking was obtained by presenting images of a single larynx to 30 ENT-professionals in a forced pairwise comparison and asking them to select the image with the best image quality. Differences in image quality between endoscopes are reliably detected by objective measurement and subjective assessment. A strong positive correlation was found between sharpness and subjective ranking (p 0.005). Visual noise and color fidelity may be relevant, but did not correlate with the subjective assessment and were probably overshadowed by the strong correlation between sharpness and subjective ranking in the data. The authors found that number of pixels on screen to display the registered image differs per type of endoscope, however more pixels do not necessarily imply a sharper image. The authors will continue their investigation of image quality metrics and their relation to diagnostic accuracy which can provide feedback on design and manufacturing optimization to the industry. (C) 2022 Society for Imaging Science and Technology.
机译:耳鼻喉科柔性内窥镜是耳鼻喉科(ENT)专业人员检查上气道的重要工具。尽管图像质量在过去十年中有了显着提高,但还没有普遍接受的方法来客观地衡量这一点。清晰度、视觉噪点和色彩保真度是可以客观测量的图像质量方面。本研究的目的是探讨这些质量指标与耳鼻喉科专业人员对图像质量的主观感知之间的关系。对6种不同软式内窥镜的图像质量进行客观和主观评价。使用Rez Checker Target Nano Matte获得客观测量值,包括清晰度(MTF50)、视觉噪点和色彩保真度(CIE Delta E 2000)。通过强制成对比较向 30 名耳鼻喉科专业人员展示单个喉部的图像,并要求他们选择具有最佳图像质量的图像,获得主观图像质量排名。通过客观测量和主观评估可靠地检测出内窥镜之间的图像质量差异。发现锐度与主观排名之间存在很强的正相关关系(p < 0.005)。视觉噪点和色彩保真度可能是相关的,但与主观评估无关,并且可能被数据中清晰度和主观排名之间的强相关性所掩盖。作者发现,屏幕上显示注册图像的像素数量因类型的内窥镜而异,但是像素越多并不一定意味着图像越清晰。作者将继续研究图像质量指标及其与诊断准确性的关系,从而为行业提供有关设计和制造优化的反馈。(C) 2022 年影像科学与技术学会。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号