Technical note: Four‐year experience with utilization of DICOM metadata analytics in clinical digital radiography practice

机译:Technical note: Four‐year experience with utilization of DICOM metadata analytics in clinical digital radiography practice

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Abstract Background Digital radiography (DR) still presents many challenges and could have complex imaging acquisition and processing patterns in a clinical practice hindering quality standardization. Purpose This technical note aims to report the 4‐year experience with utilizing a custom DICOM metadata analytics program in clinical DR at a large institution. Methods Thirty‐eight DR systems of three vendors at multiple locations were configured to automatically send clinical DICOM images to a DICOM receiver. A suite of custom MATLAB programs was established to extract and store public and private header data weekly. Specific use cases are provided for systematic image acquisition investigation, image processing harmonization, exposure index (EI) longitudinal monitoring and EI target optimization. Results For systematic acquisition investigation, an example of adult lumbar spine exam analysis was provided with statistics on manual acquisition versus the use of automatic exposure control (AEC, including AEC dose level, active cell, and backup timer), grid usage, and collimation for various projections. For processing harmonization, up to 12.6 of protocols were revealed to have processing parameter differences in an example of a mobile radiography fleet. In addition, inconsistent use of a post‐acquisition image size function was also demonstrated, which resulted in anatomy size display variations. Bimonthly monitoring of median EI values showed expected trends, including changes after an AEC dose level adjustment for adult posterior‐anterior chest imaging on a scanner system. An example of adult axillary shoulder EI target refinement was shared using the median value, eμ, based on the lognormal EI data distribution after parsing down to acquisitions with appropriate techniques. Conclusions This analytics program enables systematic analysis of image acquisition and processing details. The information provides invaluable insights into real practice patterns, which can support data‐driven quality standardization and optimization.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号