Real-time assessment of anteroposterior stability of spinal segments

机译:Real-time assessment of anteroposterior stability of spinal segments

获取原文
获取原文并翻译 | 示例
           

摘要

Abstract Purpose While anteroposterior instability of spinal segments is regarded as an important biomechanical aspect in the clinical evaluation of lumbar pathologies, the reliability of the available diagnostic tools is limited and an intraoperative method to quantify stability is lacking. The aim of this study was to develop and validate an instrument to measure the anteroposterior stability of a spinal segments in real-time.Methods Torsi of five fresh-frozen human cadavers were used for this study. After pedicle screw insertion, a specifically modified reposition tool composed with load and linear sensors was used to measure the segmental anteroposterior motion caused by 100?N anterior and posterior force during 5 loading cycles on either side of the instrumentation by two different operators. The spinal segments were then resected from the torsi and anteroposterior loading with?±?100?N was repeated in an advanced biomechanical spine testing setup as a reference measurement. The Inter-correlation coefficient (ICC) was used for validation of the “intraoperative” device. Results Inter-operator repeatability of the measurements showed an ICC of 0.93 (p?