Patient dose estimation in CT examination using dose conversion coefficient method and CT -expo software

机译:使用剂量转换系数法和CT-expo软件进行CT检查中的患者剂量估算

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The quantity of radiation delivered to patients undergoing computed tomography (CT) examination is the first step in the optimization of patient dose. The purpose of this study was to estimate and validate adult patient dose during CT examination at the Greater Accra Regional Hospital, Ridge-Accra by using conversion coefficient and CT-Expo software. Organ and effective doses for 150 patients undergoing head, chest and abdominopelvic examinations were estimated using DLP to effective dose conversion coefficient method and CT-Expo software. Mean organ dose estimated using CT-Expo software were compared with other international published data. Organ dose from this study were generally lower as compared to organ dose values from other studies. The eye lens received the highest mean organ dose of 38.81 +/- 5.01 mGy whilst the heart received the lowest mean organ dose of 5.26 +/- 2.03 mGy. No statistical differences (p <= 0.051) were observed between the median effective doses estimated from the two methods for the chest and abdominopelvic CT examinations. Significant (p = 0.001) differences were observed between the median effective dose estimated by the two techniques for the head examination. The median effective dose for DLP to effective dose conversion coefficient (k-coefficient) method and CT-Expo software for the head CT were 1.605 and 2.00 mSv respectively. The mean organ dose per each CT examination from this study were generally comparable with those from other published data.
机译:向接受计算机断层扫描 (CT) 检查的患者输送的辐射量是优化患者剂量的第一步。本研究的目的是通过使用转换系数和 CT-Expo 软件在 Ridge-Accra 大阿克拉地区医院进行 CT 检查时估计和验证成年患者剂量。使用 DLP 到有效剂量转换系数法和 CT-Expo 软件估计了 150 例接受头部、胸部和腹部盆腔检查的患者的器官和有效剂量。使用 CT-Expo 软件估计的平均器官剂量与其他国际发表的数据进行了比较。与其他研究的器官剂量值相比,本研究的器官剂量通常较低。眼睛晶状体接受的最高平均器官剂量为 38.81 +/- 5.01 mGy,而心脏接受的最低平均器官剂量为 5.26 +/- 2.03 mGy。胸部和腹部盆腔CT检查的两种方法估计的中位有效剂量之间没有观察到统计学差异(p <= 0.051)。在两种技术估计的头部检查的中位有效剂量之间观察到显着(p = 0.001)差异。头部CT的DLP到有效剂量转换系数(k系数)方法和CT-Expo软件的中位有效剂量分别为1.605和2.00 mSv。本研究每次 CT 检查的平均器官剂量通常与其他已发表数据相当。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号