Anatomical accuracy of the KODEX-EPD novel 3D mapping system of the left atrium during pulmonary vein isolation: A correlation with computer tomography imaging

机译:肺静脉隔离过程中左心房KODEX-EPD新型3D标测系统的解剖精度:与计算机断层扫描成像的相关性

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Background A novel 3D mapping system (KODEX-EPD, EPD Solutions) enables catheter localization and real-time 3D cardiac mapping. Objective To evaluate left atrium (LA) anatomical mapping accuracy created by the KODEX-EPD system during pulmonary vein isolation (PVI) compared with gold standard computed tomography (CT) images acquired from the same patients before the procedure. Methods In 15 consecutive patients who underwent PVI, 3D mapping of the LA was created on the KODEX-EPD system using the Achieve catheter. Pulmonary vein (PV), posterior wall, and appendage anatomy and diameters, were compared to the CT 3D reconstruction measured on the CARTO 3 system. Measurements were done independently by two physicians in each method. Linear correlation and agreement between CT and EPD measurements were assessed by Spearman correlation and Bland-Altman plot. Results Mean LA mapping time was 7.7 +/- 3.6 min. Very high interobserver correlation was found for both EPD and CT measurements (Spearman r = .9). High correlation (r = .75) was found between CT and EPD measurements. Bland-Altman plot method revealed that measurements assessed by EPD were slightly higher than those assessed by CT. Mean difference was 3.5 mm, p < .01. In 2 (13.5) patients each, disagreement regarding the presence of a left common PV and a right middle accessory vein anatomy was seen. Conclusion The new KODEX-EPD mapping system allows quick and accurate mapping of the LA with high correlation to CT imaging. Some differences in left common and accessory right middle vein anatomy were seen.
机译:背景 一种新型的 3D 标测系统(KODEX-EPD、EPD Solutions)可实现导管定位和实时 3D 心脏标测。目的 评估KODEX-EPD系统在肺静脉隔离(PVI)过程中产生的左心房(LA)解剖标测精度与术前相同患者采集的金标准计算机断层扫描(CT)图像的比较。方法 在连续 15 例接受 PVI 的患者中,使用 Achieve 导管在 KODEX-EPD 系统上创建 LA 的 3D 映射。将肺静脉 (PV)、后壁和附件解剖结构和直径与在 CARTO 3 系统上测量的 CT 3D 重建进行比较。每种方法的测量均由两名医生独立完成。通过Spearman相关性和Bland-Altman图评估CT和EPD测量之间的线性相关性和一致性。结果 平均LA映射时间为7.7 +/- 3.6 min。EPD 和 CT 测量均发现观察者间相关性非常高 (Spearman r = .9)。CT 和 EPD 测量之间存在高度相关性 (r = .75)。Bland-Altman图法显示,EPD评估的测量值略高于CT评估的测量值,平均差为3.5 mm,p < 0.01。在各 2 例 (13.5%) 患者中,观察到对左侧共同 PV 和右侧中副静脉解剖结构存在分歧。结论 新型KODEX-EPD标测系统可快速、准确地绘制与CT成像高度相关的LA标测。左总静脉和副右中静脉解剖结构存在一些差异。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号