Effect of spatial averaging on the amplitude ring ratio in multifocal electroretinography

机译:Effect of spatial averaging on the amplitude ring ratio in multifocal electroretinography

获取原文
获取原文并翻译 | 示例
           

摘要

Abstract Purpose The purpose of this study was to evaluate the effect of spatial averaging on the multifocal electroretinography (mfERG) amplitude ring ratios used in screening for hydroxychloroquine (HCQ) toxicity.Methods This was a retrospective review of the records of patients screened for HCQ retinopathy at the USF Eye Institute (University of South Florida) during the period of 2015–2020. Patients were tested binocularly with Diagnosys mfERG system (Diagnosys LLC, Lowell, MA). Only the records of patients referred internally were used. The effects of the lowest level (level 1, or 8%) of spatial averaging on the P1 amplitude ring ratios used for screening of HCQ maculopathy: R1/R2, R2/R5, R5/R3 and R5/R4, were evaluated.Results The records of 40 patients (4 males, 36 females) aged 54.4?±?14.1 years were selected for analysis. The use of spatial averaging had a significant effect on P1 amplitudes, and on the ring ratios and this effect was correlated with the magnitude of the amplitudes and the ratios. Spatial averaging diminished P1 amplitude significantly in ring 1 (p?