Novel use of an irrigated ablation catheter to monitor real‐time hemodynamics during ablation

机译:Novel use of an irrigated ablation catheter to monitor real‐time hemodynamics during ablation

获取原文
获取原文并翻译 | 示例

摘要

Abstract Introduction Hemodynamic decompensation during catheter ablation occurs due to prolonged procedure time and irrigant delivery directly into the cardiac chambers. Real‐time hemodynamic monitoring of patients undergoing catheter ablation procedures may identify patients at risk of decompensation; we set out to assess the feasibility of a novel, real‐time, intracardiac pressure monitoring system using a standard irrigated ablation catheter. Methods We studied 13 consecutive who underwent pressure measurement of the left atrium (LA) and left ventricle (LV) via transeptal access with a Swan Ganz (SG) catheter followed by two commercially available irrigated ablation catheters. Pressure waveform data was extracted to compare LA peak pressure, LV peak systolic pressure, LV end‐diastolic pressure, and waveform analysis. Results Comparison between the SG and ablation catheters (AblA; AblB) demonstrated that LV systolic pressure (0.61–16.8?mmHg; 1.32–18.2?mmHg), and LV end‐diastolic pressure (?3.4 to 2.8?mmHg; ?3.0 to 3.35?mmHg) were well correlated and had accepted repeatability. Ablation waveforms demonstrated an 89.9?±?6.4% correlation compared to SG waveforms. Conclusion Pressure measurements derived from an irrigated ablation catheter are accurate and reliable when compared to an SG catheter. Further studies are needed to determine how real‐time pressure monitoring can improve outcomes during ablation procedures.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号