首页> 外文期刊>Journal of cataract and refractive surgery >Comparative analysis of 2 biometers using swept-source OCT technology
【24h】

Comparative analysis of 2 biometers using swept-source OCT technology

机译:使用扫描源 OCT 技术对 2 个生物计进行比较分析

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Purpose: To report the repeatability, level of agreement, and correlation of 2 swept-source optical coherence tomography (SS-OCT) biometers, the lOLMaster 700 (biometer A) and the ANTERION (biometer B).Setting: Rothschild Foundation Hospital, Paris, France. Design: Prospective comparative case series.Methods: Consecutive patients who attended for cataract assessment had SS-OCT biometry performed with the reference, biometer A, and biometer B. Axial length, mean keratometry, central corneal thickness, anterior chamber depth, lens thickness, and corneal diameter measurements (white-to-white WTW distance) were measured. The intraoperator repeatability was calculated using analysis of variance and repeatability limits.Correlations were assessed, and the level of agreement between the biometers was represented with the Bland-Altman method.Results: The study comprised 63 patients (125 eyes). There was a statistically significant difference between the 2 biometers in all measurements (P < .05). All repeatability measurements were very high, as were the level of correlation and level of agreement between biometer A and biometer B for all parameters.Conclusions: Biometer B provided good agreement and repeatability compared with biometer A; however, all parameters were not interchangeable, particularly, the WTW measurement and the high keratometric values.
机译:目的:报告 2 个扫描源光学相干断层扫描 (SS-OCT) 生物计 lOLMaster 700(生物计 A)和 ANTERION(生物计 B)的重复性、一致性水平和相关性。地点:法国巴黎罗斯柴尔德基金会医院。设计:前瞻性比较案例系列。方法:连续参加白内障评估的患者使用参考、生物计 A 和生物计 B 进行 SS-OCT 生物测量,测量眼轴长度、平均角膜曲率、中央角膜厚度、前房深度、晶状体厚度和角膜直径测量值(白白 [WTW] 距离)。使用方差和重复性限值分析来计算操作员内部的重复性。评估相关性,并用Bland-Altman方法表示生物计之间的一致性水平。结果:该研究包括 63 名患者(125 只眼睛)。在所有测量中,2 个生物计之间存在统计学上的显着差异 (P < 0.05)。所有重复性测量都非常高,生物计 A 和生物计 B 之间所有参数的相关性和一致性水平也非常高。结论:与生物计A相比,生物计B具有较好的一致性和重复性;然而,所有参数都不可互换,特别是WTW测量值和高角膜测量值。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号