首页> 外文期刊>Journal of interventional cardiac electrophysiology: an international journal of arrhythmias and pacing >The target ablation index values for electrical isolation of the superior vena cava
【24h】

The target ablation index values for electrical isolation of the superior vena cava

机译:The target ablation index values for electrical isolation of the superior vena cava

获取原文
获取原文并翻译 | 示例
           

摘要

Abstract Purpose The ablation index (AI), developed as a radiofrequency (RF) catheter ablation composite component endpoint, which incorporates contact force (CF), time, and power in a weighted formula, has been reported to be useful for a durable pulmonary vein isolation (PVI) to treat atrial fibrillation (AF). No study has reported the target AI value for the SVC isolation (SVCI). In this study, we aimed to investigate the target AI for the SVCI.Methods Thirty-six AF patients who underwent an initial SVCI were enrolled. Ablation was performed at 556 points. The sites where dormant conduction was induced or additional ablation was needed were defined as touch up sites (n?=?36). We compared the energy deliver time, power, generator impedance (GI) drop, local bipolar voltage, contact force (CF), force–time integral (FTI), and AI between the touch up sites and the no touch up sites (n?=?520).Results The FTI and AI were significantly lower at the touch up sites (touch up sites vs. no touch up sites; FTI, 126.5 [99.3–208.8] vs. 244 [184–340.8], p?