Purifying air over the operating field with a new mobile laminar airflow device to reduce the possibility of airborne contamination

机译:使用新型移动层流气流装置净化操作区域内的空气,以减少空气污染的可能性

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Purpose: To investigate the effectiveness of a mobile laminar airflow (LAF) device designed to reduce both airborne particles and lint fibers within the sterile field during cataract surgery. Setting: Cincinnati Eye Institute, Cincinnati, Ohio. Design: Prospective case series. Methods: The number of airborne particles sized 0.5 mu m, 1.0 mu m, and 1.5 mu m were measured at different locations and times during 116 routine phacoemulsification procedures with and without the use of a mobile LAF device. In a companion study, the presence of lint fibers in 99 eyes undergoing cataract surgery with a mobile LAF device was compared with 50 eyes in the control group. Lint fibers were differentiated as either falling onto the ocular surface or being carried into the sterile field by an instrument. Results: A statistically significant reduction (P < .0005) occurred in all 3 particle sizes measurements when the LAF device was used: 79.0 (93.6 +/- 16.1 vs 445.3 +/- 30.9), 81.7 (11.30 +/- 1.98 vs 61.85 +/- 5.34), and 90.8 (0.241 +/- 0.056 vs 2.624 +/- 0.362) in the 0.5 mu m, 1.0 mu m and 5.0 mu m particles, respectively. Lint fibers were identified in 18 (9/50) of eyes in the control group and 16.16 (16/99) of eyes in the LAF group. Although the number of lint fibers carried into the sterile field was similar in each group, the incidence of lint fibers falling onto the sterile field was reduced from 6 (3/50) to 0 (0/99) when the LAF was used (P = .014). Conclusions: The mobile LAF device was highly effective in reducing the number of particulate matter and lint fibers within the sterile surgical field when used during cataract surgery. J Cataract Refract Surg 2021; 47:1327-1332 Copyright (c) 2021 Published by Wolters Kluwer on behalf of ASCRS and ESCRS
机译:目的:研究一种移动层流 (LAF) 装置的有效性,该装置旨在减少白内障手术期间无菌区域内的空气传播颗粒和棉绒纤维。地点:俄亥俄州辛辛那提市辛辛那提眼科研究所。设计:前瞻性案例系列。方法:在116例常规超声乳化术中,使用和不使用移动LAF装置,测量大小为0.5 μ m、1.0 μ m和1.5 μ m的空气传播颗粒的数量。在一项伴随研究中,将 99 只使用移动 LAF 设备进行白内障手术的眼睛与对照组的 50 只眼睛中棉绒纤维的存在进行了比较。棉绒纤维被区分为落在眼表或被仪器带入无菌区域。结果:当使用LAF设备时,所有3种粒径测量均发生统计学显着降低(P < .0005):0.5 μ m、1.0 μ m和5.0 μ m颗粒分别为79.0%(93.6 +/- 16.1 vs 445.3 +/- 30.9)、81.7%(11.30 +/- 1.98 vs 61.85 +/- 5.34)和90.8%(0.241 +/- 0.056 vs 2.624 +/- 0.362)。对照组18%(9/50)的眼睛和16.16%(16/99)的眼睛分别检出绒毛纤维。尽管每组携带到无菌区域的棉绒纤维数量相似,但当使用LAF时,皮棉纤维落入无菌区域的发生率从6%(3/50)降低到0%(0/99)(P = .014)。结论:在白内障手术中使用移动式LAF装置时,可有效减少无菌手术区域内的颗粒物和棉绒纤维数量。J 白内障屈光外科杂志 2021;47:1327-1332 版权所有 (c) 2021 由 Wolters Kluwer 代表 ASCRS 和 ESCRS 发布

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号