首页> 外文期刊>chemistryselect >Interference Free Simultaneous Detection of Dihydroxy Benzene Isomers at Cost-effective and Reliable Celestine Blue Modified Glassy Carbon Electrode
【24h】

Interference Free Simultaneous Detection of Dihydroxy Benzene Isomers at Cost-effective and Reliable Celestine Blue Modified Glassy Carbon Electrode

机译:Interference Free Simultaneous Detection of Dihydroxy Benzene Isomers at Cost-effective and Reliable Celestine Blue Modified Glassy Carbon Electrode

获取原文
获取原文并翻译 | 示例
           

摘要

A successful cyclic voltammetric electropolymerisation and electrochemical characterization of celestine blue (CB) was achieved on glassy carbon electrode (GCE). The fabricated electrode was then used for the electroanalysis of hydroquinone (HQ) and catechol (CC). Due to the improved electrocatalytic activity of the fabricated electrode, an interference free simultaneous detection of CC by HQ was observed. The scan rate studies suggest the diffusion-controlled kinetics phenomenon at the modified electrode. Concentration studies, yielding a limit of detection (LoD) of 0.0491 mu M and 0.0456 mu M for CC and HQ respectively in the first linear range of 0 to 40 mu M, limit of detection of 0.0621 mu M and 0.0638 mu M for CC and HQ respectively in the second linear range of 50 to 120 mu M, which were relatively lower as compared to previous reports was achieved by using differential pulse voltammetry (DPV). In order to evaluate the analytical performance of the CB/GCE, tap water sample analysis was carried out and good recovery results were obtained.
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有