Muscle usage and workload assessment of cardiac ablation procedure with the use of a novel catheter torque tool in a pediatric simulator

机译:Muscle usage and workload assessment of cardiac ablation procedure with the use of a novel catheter torque tool in a pediatric simulator

获取原文
获取原文并翻译 | 示例
           

摘要

Background Cardiac ablation catheters are small in diameter and pose ergonomic challenges that can affect catheter stability. Significant finger dexterity and strength are necessary to maneuver them safely. We evaluated a novel torque tool to reduce muscle activation when manipulating catheters and improve perceived workload of ablation tasks. The objective was to evaluate measurable success, user perception of workload, and muscle usage when completing a simulated ablation task with and without the use of a catheter torque tool. Methods Cardiology attendings and fellows were fitted with surface electromyographic (EMG) sensors on 6 key muscle groups in the left hand and forearm. A standard ablation catheter was inserted into a pediatric cardiac ablation simulator and subjects navigated the catheter tip to 6 specific electrophysiologic targets, including a 1-min simulated radiofrequency ablation lesion. Time to complete the task, number of attempts required to complete the lesion, and EMG activity normalized to percentage of maximum voluntary contraction were collected throughout the task. The task was completed 4 times, twice with and twice without the torque tool, in semi-randomized order. A NASA Task Load Index survey was completed by the participant at the conclusion of each task. Results Time to complete the task and number of attempts to create a lesion were not altered by the tool. Subjectively, participants reported a significant decrease in physical demand, effort, and frustration, and a significant increase in performance. Muscle activation was decreased in 4 of 6 muscle groups. Conclusion The catheter torque tool may improve the perceived workload of cardiac ablation procedures and reduce muscle fatigue caused by manipulating catheters. This may result in improved catheter stability and increased procedural safety.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号