首页> 外文期刊>Surgical innovation >New Chest Drainage Unit with Integrated CO2-Detector Demonstrates False Air Leak
【24h】

New Chest Drainage Unit with Integrated CO2-Detector Demonstrates False Air Leak

机译:New Chest Drainage Unit with Integrated CO2-Detector Demonstrates False Air Leak

获取原文
获取原文并翻译 | 示例
           

摘要

Background Postoperative observed air leakage does not always originate from parenchymal defects but may arise from defects in the chest drainage unit, connections or reverse airflow in water seals. We investigated such false air leakage using a new chest drainage unit with a built-in CO2-detector and an electronic chest drainage unit. Methods Two types of chest drainage units were tested in a simple porcine model: A well-known electronic chest drainage unit and a new chest drainage unit with integrated CO2-detector. We created a setup of true air leakage-a parenchymal lesion, and false air leakage-allowing air to flow into the thoracic cavity alongside the chest drain. Results We demonstrated that the new chest drainage unit with a built-in CO2-detector can distinguish between experimentally induced true air leakage and false air leakage. Conclusion Available chest drainage systems do not allow direct assessment of true or false air leakage, which may increase chest drain duration unnecessarily. The integration of a CO2-sensitive color indicator into a chest drainage unit allows simple distinction between false air leak and true air leak, which may improve postoperative management.