The thermal field of radiofrequency probes at chondroplasty settings.

机译:射频探头的温度场软骨成形术设置。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: The purpose of this experimental study was to evaluate the thermal field produced with monopolar and bipolar radiofrequency probes located at increasing heights from the target treatment site. TYPE OF STUDY: Experimental study. METHODS: Two bipolar (ACD-50, ArthroCare, Sunnyvale, CA; VAPR-TC, Mitec Surgical Products, Westwood, MA) and one monopolar (Vulcan TAC-C II, Oratec Interventions, Menlo Park, CA) radiofrequency probes were placed in a screw-driven stage that allowed for 0.25-mm incremental height-position changes. ArthroCare ACD-50 was evaluated at settings 2 and 8. The Mitek VAPR-TC was evaluated at setting 65 degrees C and V2-20 desiccation mode. The Oratec Vulcan TAC-C II system was evaluated at preset 2, power 15, set temperature 70 degrees C. The RF-probes were evaluated at variable distances from 0 to 2 mm. A fluoroptic thermometer was used to evaluate the temperatures within a room temperature 0.9% normal saline arthroscopic simulation chamber. RESULTS: The ACD-50 setting 2 at 0 mm was 89.1 degrees C; at 0.5 mm, 71.2 degrees C; and at 2 mm it was 37.3 degrees C. The ACD-50 setting 8 at 0 mm was 87.3 degrees C; at 0.5 mm, 42.5 degrees C; and at 2 mm, 33.6 degrees C. The Mitek VAPR-TC at 0 mm was 53.1 degrees C; at 0.5 mm, 42.5 degrees C; and at 2 mm, 26.8 degrees C. The Oratec Vulcan at 0 mm was 73.9 degrees C; at 0.5 mm, 60.3 degrees C; and at 2 mm, 29.0 degrees C. Each of these radiofrequency systems produced characteristic thermal fields at these settings. CONCLUSIONS: The temperature decreases with increasing distance characteristically for each radiofrequency probe. This is clinically important because altering radiofrequency probe location may cause large variations in articular cartilage thermal exposure.
机译:目的:本实验研究的目的是评估产生的温度场单极和双极射频探头位于增加高度的目标处理网站。研究。加州桑尼维尔;韦斯特伍德,MA)和一个单极(火神TAC-C二世,Oratec干预,门洛帕克,CA)射频探头放置在一个研究阶段,允许为0.25毫米增量高度位置的变化。ACD-50评估在设置2和8。Mitek VAPR-TC在设置65度评估C和V2-20干燥模式。TAC-C II系统评估在预设2,力量15日,设定温度70度c RF-probes在变量的距离从0到2进行评估吗毫米。fluoroptic温度计用于评估温度在室温下0.9%生理盐水模拟关节镜室。结果:ACD-50设置2 0毫米是89.1度;这是37.3度c ACD-50设置8 0mm是87.3摄氏度;在2毫米,33.6度c Mitek VAPR-TC0毫米53.1摄氏度;C;在0毫米73.9摄氏度;度;这些射频系统产生在这些设置热特征字段。结论:温度降低典型的为每个增加距离射频探头。重要,因为改变射频探头位置可能导致大关节的变化软骨热接触。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号