Ascending Defibrillation Waveform Significantly Reduces Myocardial Morphological Damage and Injury Current

机译:显著提升除颤波形减少心肌形态学损伤和损伤电流

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVES This study tested the hypothesis that a biphasic defibrillation waveform with an ascending first phase (ASC) causes less myocardial damage by pathology and injury current than a standard biphasic truncated exponential (BTE) waveform in a swine model.BACKGROUND Although lifesaving, defibrillation shocks have significant iatrogenic effects that reduce their benefit for patient survival.METHODS An ASC waveform with an 8-ms linear ramp followed by an additional positive 0.5-ms decaying portion with amplitudes of 20 J (ASC 20J) and 25 J (ASC 25J) was used. The control was a 25-J BTE conventional waveform (BTE 25J)
机译:目的本研究的假说进行测试与一个两相的除颤波形提升第一阶段(ASC)导致更少心肌损伤的病理和伤害比标准的两相的截断指数(耳背式)波形在猪模型。尽管救生,去颤冲击重要的,减少医源性影响对病人的生存受益。波形与一个线性斜坡紧随其后的是一个8毫秒另外积极的0.5 - ms腐烂的部分振幅20 J (ASC 20)和25 J (ASC 25 J)是使用。波形(耳背式25 j)