首页> 外文期刊>JACC. Clinical electrophysiology. >Is the Best (Waveform) the Enemy of the Good (Waveform)? A Defibrillation Challenge
【24h】

Is the Best (Waveform) the Enemy of the Good (Waveform)? A Defibrillation Challenge

机译:是最好的(波形)的敌人好吗(波形)?

获取原文
获取原文并翻译 | 示例
           

摘要

Since the demonstration of successful clinical defibrillation by electrical shock in 1947 (1), the evolution of defibrillators has focused on the pursuit of defibrillator waveforms that can reliably and reproducibly terminate ventricular fibrillation without incurring myocardial damage. The waveform of an electrical shock is defined as the temporal pattern of its amplitude measured by voltage or current. After a long period of monophasic waveform configurations, a more efficient biphasic waveform was introduced. All current implantable cardioverter-defibrillators (ICD) use biphasic truncated exponential (BTE) waveforms to defibrillate. The use of these exponential decaying morphologies is based on the simplicity and reliability of modern capacitors as the main component of energy delivery, which in turn, makes it possible to generate the current in relatively small implantable devices.
机译:因为成功的临床的示范电击除颤的1947 (1),心脏除颤器一直专注于的进化去纤颤器波形可以的追求可靠的和可重复终止室纤维性颤动不引起心肌损伤。电击的波形被定义为时间模式的振幅来衡量电压或电流。单相波形配置,更多介绍了高效的两相的波形。当前植入型心律转复除颤器(ICD)使用两相的截断指数(耳背式)除纤颤的波形。基于指数衰减的形态现代电容器的简单性和可靠性能源输送的重要组成部分,反过来,可以生成电流相对较小的植入式设备。