MR-based thermometry of laser induced thermotherapy: Temperature accuracy and temporal resolution in vitro at 0.2 and 1.5 T magnetic field strengths

机译:MR-based温度测量的激光诱导温热疗法:温度准确性和时间分辨率在0.2和1.5 T体外磁场磁场强度

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose To evaluate MR-thermometry using fast MR sequences for laser induced interstitial thermotherapy (LITT) at 0.2 and 1.5 T systems. Methods & Materials In-vitro experiments were performed using Agarose gel mixture and lobes of porcine liver. MR-thermometry was performed by means of longitudinal relaxation time (T1) and proton resonance frequency shift (PRF) methods under acquisition of amplitude and phase shift images. Four different sequences were used for T1 thermometry: A gradient-echo (GRE), a True Fast Imaging with Steady Precession (TRUFI), a Saturation Recovery Turbo-FLASH (SRTF), and an Inversion Recovery Turbo-FLASH (IRTF) sequence (FLASHaFast Low Angle Shot). PRF was measured with four sequences: Two fast-spoiled GRE sequences (one as WIP sequence), a Turbo-FLASH (TFL) sequence (WIP sequence), and a multiecho-TrueFISP sequence. Temperature was controlled and verified using a fiber-optic Luxtron device. The temperature was correlated with the MR measurement. Results All sequences showed a good linear correlation R 2 = 0.97-0.99 between the measured temperature and the MR-thermometry measurements. The only exception was the TRUFI sequence in the Agarose phantom that showed a non-linear calibration curve R 2 = 0.39-0.67. At 1.5 T, the Agarose experiments revealed similar temperature accuracies of 4-6°C for all sequences excluding TRUFI. During experiments with the liver, the PRF sequences showed better performance than the T1, with accuracies of 5-12°C, contrary to the T1 sequences at 14-18°C. The accuracy of the Siemens PRF-FLASH sequence was 5.1°C. At 0.2 T, the Agarose experiments provided the highest accuracy of 3.3°C for PRF measurement. At the liver experiments the T1 sequences SRTF and FLASH revealed the best accuracies at 6.4 and 7.0°C. Conclusion The accuracy and speed of MR temperature measurements are sufficient for controlling the temperature-based tumor destruction. For 0.2 T systems SRTF and FLASH sequences are recommended. For 1.5 T systems SRTF and FLASH are the most accurate.
机译:目的评价MR-thermometry使用快速先生序列激光诱导间质温热疗法(LITT) 0.2和1.5 T系统。方法,使用琼脂糖凝胶混合物和叶的执行猪的肝脏。的纵向弛豫时间(T1)和质子共振频移(脉冲)方法在收购振幅和相移图像。温度测量:gradient-echo (GRE),一个真正的快成像与稳定的旋进(TRUFI)饱和恢复Turbo-FLASH (SRTF),和一个反转恢复Turbo-FLASH (IRTF)那样,序列(FLASHaFast低角度拍摄)。有四个序列:两个fast-spoiled GRETurbo-FLASH序列(一个在制品序列)(伦敦交通局)序列(在制品序列),multiecho-TrueFISP序列。使用光纤控制和验证Luxtron设备。先生与测量。呈良好的线性相关(R = 0.97至-0.99点测量的温度和之间MR-thermometry测量。TRUFI序列在琼脂糖幻影显示一个非线性校准曲线R 2 =0.39 - -0.67。显示类似的温度精度4 - 6°C所有序列TRUFI除外。实验与肝脏、PRF序列比T1,表现出更好的性能精度的5 - 12°C,相反T1序列在14 - 18°C。PRF-FLASH序列为5.1°C。琼脂糖的实验提供了最高的精度3.3°C的脉冲重复频率测量。实验SRTF T1序列和FLASH发现最好的精度在6.4和7.0°C。结论的准确性和速度温度测量是充分的控制temperature-based肿瘤破坏。序列是推荐。和闪电是最准确的。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号