Safety of a continuous glucose monitoring device during hyperbaric exposure.

机译:连续葡萄糖监测设备的安全在高压曝光。

获取原文
获取原文并翻译 | 示例
           

摘要

Hyperbaric oxygen therapy has been demonstrated to lower blood glucose levels in patients with diabetes. Continuous glucose monitoring (CGM) allows glucose monitoring in real time. Battery-operated CGM transmitters have yet to be formally tested and given safety approval for use in a hyperbaric environment. We evaluated and tested commercially available Dexcom? G6 CGM transmitters under hyperbaric conditions. Each transmitter contains a 3V, 130-mAh (0.39 Wh) lithium manganese dioxide battery (IEC CR1632) and circuit board that are fully encapsulated in epoxy. Each transmitter is pressurized to 90 pounds per square inch (psi) in an autoclave at 40°C for up to 72 hours during manufacturing to ensure that all enclosed air spaces are eliminated from the epoxy. We compared the CGM components against section 14.2.9.3.17.5 of the 2018 National Fire Protection Association 99 (NFPA 99) Health Care Facilities Code requirements. Six CGM transmitters attached to estimated glucose value generators (EGVGs) underwent 11 pressurization cycles to 45 feet of seawater (fsw). All transmitters were returned to the manufacturer to assess post-exposure structural integrity. G6 sensors, which contain no electrical components or compressible air spaces, do not pose a risk in the hyperbaric environment. There was no observed change in preset EGVG readings during hyperbaric exposures. Post-exposure testing revealed no structural compromise after repeated hyperbaric exposures. The CGM transmitter meets section 14.2.9.3.17.5 of the 2018 NFPA 99 requirements for battery-operated devices allowed for use in a hyperbaric environment. This analysis revealed no significant safety concerns with subjecting Dexcom G6 CGM transmitters to hyperbaric environments.
机译:高压氧疗法已被证明降低患者的血糖水平糖尿病使血糖实时监控。电池驱动的CGM发射器尚未正式测试和安全批准使用在高压环境中。测试商用? Dexcom公司将在高压条件下发射器。发射机包含3 v, 130 - mah (0.39 Wh)锂二氧化锰电池(IEC CR1632)和电路板完全封装在环氧树脂。磅每平方英寸(psi)的高压蒸汽40°C 72小时在制造业确保所有封闭的空气空间从环氧消除。对部分14.2.9.3.17.5组件2018年全国防火协会99年(NFPA 99)卫生保健设施代码要求。估计葡萄糖值生成器(EGVGs)经历了11增压周期45英尺海水(fsw)。制造商评估接触后结构的完整性。没有电器元件或可压缩空气空间,在高压不构成威胁环境。在高压暴露预设EGVG读数。接触后测试显示没有结构反复高压暴露后妥协。14.2.9.3.17.5 CGM发射机满足部分2018年NFPA 99要求电池驱动的设备允许使用高压的环境。对重大安全隐患G6 CGM Dexcom公司将发射机高压环境。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号