首页> 外文期刊>Clinical & experimental optometry: journal of the Australian Optometrical Association >On repeated corneal applanation with the Goldmann and two non-contact tonometers.
【24h】

On repeated corneal applanation with the Goldmann and two non-contact tonometers.

机译:在重复角膜Goldmann扁平和两个非接触式眼压计。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Several authors believe it is necessary to randomise the order in which contact and non-contact tonometers are used in comparison studies. This study was carried out to investigate the effect of repeated applanation on the measured intraocular pressure. METHODS: One set of measurements per session was made on each of three sessions (one session per day) with the Goldmann and two non-contact tonometers (Topcon CT80 and Keeler Pulsair EasyEye), in a pre-determined order, on one randomly selected eye of 120 subjects randomised to one of two groups. For session one, only the non-contact tonometers were used to assess the intraocular pressure of both groups. For session two, either non-contact or Goldmann tonometry was performed first and this order was reversed for session 3. Average intraocular pressures were compared between sessions to determine the presence or absence of effects on the intraocular pressure caused by prior repeated applanation with the Goldmann or either one (or both) of the non-contact tonometers. RESULTS: Prior applanation with a non-contact tonometer did not cause a significant (p > 0.05) reduction of the mean pressure measured with either non-contact tonometer. The mean pressure was slightly but significantly (p 0.05) between the pressures measured with the Goldmann tonometer prior to and subsequent to non-contact tonometry, in both subject groups. CONCLUSION: A small but statistically significant reduction in the intraocular pressure was found following applanation with the Goldmann tonometer but not with either one of two non-contact tonometers.
机译:背景:一些作者认为需要随机选择的顺序接触和非接触式眼压计是用于比较研究。重复的扁平的影响进行调查测量眼压。每个组测量每个会话了三个会话(每天一次)Goldmann和两个非接触式眼压计(TopconCT80和基勒Pulsair EasyEye)预先确定的顺序,随机选择120名受试者随机两种组。眼压计被用来评估眼内两组的压力。非接触式或Goldmann张力测定法第一,这个订单被撤销会话3。平均眼内压力进行了比较确定存在或之间的会话缺乏对眼压的影响由于之前重复的扁平Goldmann或一个(或两个)非接触式眼压计。用非接触式眼压计没有扁平引起显著的降低(p > 0.05)平均压力与非接触式测量眼压计。显著减少(p 0.05)与之前Goldmann眼压计测量随后的非接触式张力测定法,在这两个学科组。统计上显著的减少眼压后被发现扁平Goldmann眼压计但不是用两种非接触式眼压计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号