The uncertainty in the activity estimate from a lung count due to the variability in chest wall thickness profile.

机译:从一个活动的不确定性估计肺数由于胸壁的可变性厚度。

获取原文
获取原文并翻译 | 示例
           

摘要

Calibration of a lung counter requires the use of a realistic torso phantom. The depth profile of both torso phantoms' (LLNL and JAERI) chest plate covers is fixed and assumed to be equivalent to a person's chest wall; however, ultrasound measurements of humans have shown this to be an approximation. When the depth profile of a calibration phantom is different from that of a subject, then a systematic uncertainty will be introduced into the activity estimate. Monte Carlo simulation has shown that changes in the depth profile of the chest wall thickness affect the counting efficiency. Ultrasound measurements have suggested that the coefficient of variation in the depth profile of the chest wall thickness lies between 13% and 26% for male workers; therefore, the added uncertainty to an activity estimate will be an over or underestimate of about a factor of 1.07 resulting from the different depth profile. The factor will be somewhat higher for females, probably about 1.2 at the extreme. These additional uncertainties resulting from depth profile differences are small compared with other uncertainties commonly encountered in lung counting: detector positioning, deposition patterns of the activity, measurement of the chest wall thickness, etc.
机译:肺计数器需要使用的校准一个现实的躯干幻影。躯干幻影”(LLNL和JAERI)胸部板封面是固定的,认为是相当于一个人的胸部;人类已经证明这是一个测量近似。校准幻影是不同的主题,然后将系统的不确定性引入活动估计。卡洛模拟表明变化胸壁厚度影响的深度剖面计算效率。表明,变异系数深度剖面的胸壁厚度介于男性工人的13%和26%;因此,一个活动的增加不确定性估计将超过或低估的关于1.07的一个因素不同的深度剖面。稍高一些的女性,大约1.2在极端。造成深度剖面的差异相比与其他一般的不确定性中遇到肺计数:探测器定位、沉积模式的活动,胸壁厚度测量等。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号