The second international in-vivo monitoring intercomparison program for whole body counting facilities by Canadian and United States agencies.

机译:第二国际已监测相互比对项目的整个身体计数设备由加拿大和美国机构。

获取原文
获取原文并翻译 | 示例
           

摘要

The Canadian National Calibration Reference Centre for In-Vivo Monitoring and the United States Department of Energy collaborated to offer a second international in vivo intercomparison program to whole body counting facilities in 1996. This program used a Reference Female phantom shell filled with radioactive tissue-substitute polyurethane to simulate a uniform fission-product distribution in soft tissues. The nuclides used were 137Cs and 60Co. The phantom also contained 40K homogeneously distributed in an amount similar to a Reference Female to produce a representative Compton background in the resulting spectra. Participants were asked to identify the nuclides and report activities for all except 40K. They were also asked to measure the precision of counting and supply the MDA for 137Cs and 60Co. The bias results were in the range of -30% to +80% with most facilities falling inside the range of -25% to +50% (Canadian and U.S. acceptable performance criteria). Results indicated that there was no measurable size dependency for this phantom. All reported precisions were less than 5% but NaI detector based systems seemed to have a systematic uncertainty in addition to Poisson variability. Contrarily, this was not found for Ge detector based systems. MDA data was scattered (14-3,500 Bq for 137Cs and 9-460 Bq for 60Co) and only suggested that lengthening the counting time improves MDA.
机译:加拿大国家校准参考中心体内监测和美国能源部合作提供第二国际体内相互比较项目的整个身体计数设施1996. 幻影壳牌充满放射性tissue-substitute聚氨酯来模拟统一的裂变产物分布在软组织。幻影也包含40 k均匀分布在一个类似于一个参考量女性产生康普顿代表由此产生的光谱背景。被要求识别核素和报告吗活动除了40 k。要求计算和测量精度供应MDA 137 cs和60 co。结果在-30% + 80%的范围大多数设施的范围内下降-25%+ 50%(加拿大和美国可接受的性能标准)。可测量的大小依赖这个幽灵。报告精度小于5%,但奈基于探测器的系统似乎有一个系统的不确定性除了泊松可变性。基于通用检测器的系统。(14 - 3500 Bq 137 cs和9 - 460 Bq 60 co)只建议延长计算时间提高了MDA。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号