首页> 外文期刊>Health Physics: Official Journal of the Health Physics Society >Estimation of the effective dose when protective aprons are used in medical procedures: a theoretical evaluation of several methods.
【24h】

Estimation of the effective dose when protective aprons are used in medical procedures: a theoretical evaluation of several methods.

机译:保护时估计的有效剂量围裙用于医疗程序:一个理论评价的几种方法。

获取原文
获取原文并翻译 | 示例
           

摘要

The use of the personal dose equivalent Hp(10), as measured by one or more dosimeters, in estimating the effective dose equivalent H(E) and the effective dose E was examined for situations in which a protective apron is worn by the monitored person during medical procedures. The photon energy range considered was between 0.03-1.0 MeV. Several methods recommended in the technical literature for this purpose were assessed and their ability to provide reasonable estimates for H(E) and E were compared. The assessments were theoretical and used Monte Carlo transport methods and an anthropomorphic phantom to calculate H(E), E, and Hp(10). The results showed that all of the recommended methods, using either one or more dosimeters, were applicable to this situation but that most gave good results only within limited photon energy ranges, outside of which they either considerably over-or under-estimated the doses. Some provided good estimates over the entire energy range considered.
机译:使用的个人剂量当量惠普(10)由一个或多个测试仪,在估算有效剂量当量H (E)和有效剂量E检查情况一个防护围裙穿的是监控吗人在医疗过程。0.03 - -1.0兆电子伏之间的能量范围考虑。推荐几种方法技术这个目的是评估和文学他们能够提供合理的估计H (E)和E进行了比较。理论和蒙特卡罗运输使用方法和一个拟人化的幻影计算H (E)、E和惠普(10)。所有的推荐方法,使用一个或多个测试仪,适用于此但是,大多数只给了好的结果有限光子能量范围内,外的他们要么大大)低估了剂量。估计在整个能量范围考虑。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号