The paper presents steps taken to extend the frequency range of the thermal AC-DC voltage transfer maintained at SUT up to 10 MHz. The properties of the measurement setup are described after replacing the Fluke 5700A calibrator used as AC voltage source up to 1 MHz with the Fluke 6071A RF signal generator. Next the results of the consistency checking of the AC-DC transfer difference measurements performed with the new measurement setup is presented. The results of verification of the correctness of the calculated and measured AC-DC transfer differences of the two calculable thermal AC voltage standards are presented in 1-10 MHz frequency range. The inconsistency of the measured AC-DC transfer difference is below 55 μV/V at 10 MHz, when comparing PMJTCs with SJTC and below 25 μV/V when comparing PMJTCs among themselves. The AC-DC transfer difference of the SUT calculable AC voltage standard was calculated from the modified mathematical model for frequencies up to 10 MHz. Direct comparison of the 3 V calculable AC voltage standard with the 5 V one confirmed correctness of the results obtained from the mathematical model.
机译:介绍步骤扩展频率范围的热交直流电压将保持在SUT 10 MHz。测量的属性设置在5700年取代侥幸使用的校准器作为交流电压源1 MHz的侥幸6071年,一个射频信号发生器。交直流电力传输的一致性检查不同测量与新执行测量设置。验证计算的正确性并测量交直流转移的差异两个计算热交流电压标准在1 - 10 MHz频率范围。不一致的测量交直流转移区别是低于55μV / V 10 MHz,什么时候比较PMJTCs SJTC和低于25μV / V比较PMJTCs。转移SUT的差异计算AC从修改后的电压标准计算数学模型频率10 MHz。3 V的直接比较计算AC与5 V电压标准确认获得的结果的正确性数学模型。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号