Assessment of Patient Radiation Dose during Conventional Diagnostic X-ray Examinations in Three Public Hospitals in Northern Jordan using TLDs

机译:评估患者的辐射剂量传统的诊断x射线检查三个在约旦北部公立医院使用顶级域名

获取原文
获取原文并翻译 | 示例
           

摘要

This study aims to measure entrance surface doses during routine chest and abdomen x-ray examinations of adult and child patients. Radiation dose measurements were performed using thermoluminescent dosimeters TLD-100s in three major public hospitals in northern Jordan on a total of 100 patients. Wide variations in entrance surface doses were observed within and between hospitals, which might be attributed to significant variations of the selected exposure parameters. For adult patients, the results have shown that the majority of entrance surface dose values from both chest and abdomen examinations were within recommended values of diagnostic reference levels. For child patients, the mean entrance surface dose from chest examinations in three age groups were 0.131 mGy (0-1 y), 0.136 mGy (1-5 y), and 0.191 mGy (5-10 y). These values were considered relatively high compared to the European reference levels and published results in the literature. However, for abdomen examinations, entrance surface dose values were relatively lower than European reference levels. Patient effective doses were estimated using a PCXMC 2.0 Monte Carlo program. The results for both adults and children were found to be relatively lower than the values reported by international publications. Due to the wide variations of entrance surface dose and the higher radiation doses delivered to child patients, this study recommends implementing a quality assurance program in such hospitals to achieve optimization between good image quality and minimum dose according to the as low as reasonably achievable principle. Moreover, the results of this work will provide a useful base for establishing local diagnostic reference levels for chest and abdomen examinations in Jordan.
机译:本研究旨在衡量入口表面剂量在常规胸部和腹部x光考试的成人和儿童患者。辐射剂量测量进行使用热释光的指示剂,tld - 100在三个主要的公立医院在约旦北部共有100名患者。入口和内表面剂量观察医院之间,这可能归因于选择的曝光的重要变化参数。表明,绝大多数的入口表面剂量胸部和腹部检查值在推荐值的诊断吗参考水平。从胸部检查入口表面剂量三个年龄组0.131 mGy (0 - 1 y) 0.136mGy (1 - 5 y)和0.191 mGy (5 - y)这些值被认为是相对较高的比欧洲参考水平并公布结果在文献中。考试,入学表面剂量值比欧洲参考水平相对较低。病人有效剂量估计使用2.0 PCXMC蒙特卡罗程序。两个成人和儿童被发现比报道的值相对较低国际出版物。入口表面剂量的变化高辐射剂量交付给孩子患者,这项研究建议实施质量保证程序在这样的医院之间实现优化良好的图像质量和最小剂量根据低合理可行的原则。这项工作的结果将提供一个有用的基础建立本地诊断参考胸部和腹部考试的水平乔丹。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号