首页> 外文期刊>Arthroscopy: the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association >Increased Load to Failure in Biceps Tenodesis With All-Suture Suture Anchor Compared With Interference Screw: A Cadaveric Biomechanical Study
【24h】

Increased Load to Failure in Biceps Tenodesis With All-Suture Suture Anchor Compared With Interference Screw: A Cadaveric Biomechanical Study

机译:增加在肱二头肌肌腱固定术与加载失败All-Suture缝合锚相比干扰螺钉:尸体生物力学研究

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To compare the biomechanical characteristics of a single radially expanding all-suture anchor with an interference screw for open subpectoral long head of biceps tendon (LHBT) tenodesis. Methods: Eighteen fresh-frozen matched-pair human cadaveric humeri were used for this biomechanical study. The matched pair humeri were randomly assigned into 2 experimental biceps tenodesis groups: conventional interference screw (CIS) or all-suture suture anchor (ASSA). Open subpectoral LHBT tenodesis was then performed and biomechanical testing was performed using a servohydraulic test frame. A preload of 5 N was applied for 2 minutes before cyclic loading. Displacement was recorded at cycle 300 (D300) and cycle 500 (D500) and at ultimate failure. Data recorded included displacement, load to failure, displacement at failure. Paired t test was used for analysis. Results: Decreased displacement was observed for the CIS group at D300 (1.67 +/- 0.57 mm vs 3.35 +/- 2.24 mm; P = .04), D500 (2.00 +/- 0.76 mm vs 3.87 +/- 2.20 mm; P = .04), and at failure (5.17 +/- 3.05 mm vs 10.76 +/- 2.66 mm; P < .001). Load to failure was lower in CIS specimens (170 +/- 24.5 N vs 217.8 +/- 51.54 N; P = .02). Failure in each case was tendon pullout for all CIS specimens; in ASSA 6 specimens failed as the suture pulled through the tendon, 2 specimens failed by suture breakage. No difference in stiffness was observed between the 2 groups (CIS = 98.33 +/- 22.98 N/m vs ASSA = 75.94 +/- 44.83 N/m; P = .20). Conclusions: Our study found that open subpectoral biceps tenodesis performed with an ASSA construct results in increased load to failure as compared with CIS. However, the CIS did demonstrate decreased displacement as compared to ASSA in this cadaveric biomechanical study. Clinical Relevance: ASSA and CIS at time zero provide fixation as indicated by the provider intraoperatively for LHBT tenodesis. ASSA, however, does remove less cortical bone than does CIS and therefore produces a smaller stress riser in the proximal humerus. Further testing as to the integrity of ASSA is warranted to determine the integrity of the tenodesis with cyclical loading.
机译:目的:比较生物力学一个径向扩张的特征all-suture锚的干扰螺钉打开我们长期的肱二头肌肌腱(LHBT)腱固定术。人类尸体肱骨用于匹配配对这生物力学的研究。被随机分配到2实验二头肌腱固定术组:常规螺钉干涉(CIS)或all-suture缝合锚(集团)。那时我们LHBT腱固定术和执行生物力学测试进行使用伺服液压控制的测试框架。申请前2分钟循环加载。位移被记录在300 (D300)和周期周期500 (D500)和最终失败。记录包括位移、加载失败,位移在失败。进行分析。观察D300 CIS小组(1.67 + / - 0.57毫米和3.35 + / - 2.24毫米;0.76毫米和3.87 + / - 2.20毫米;失败(5.17 + / - 3.05毫米和10.76 + / - 2.66毫米;<措施)。标本(170 + / - 24.5 vs 217.8 + / - 51.54 N;= .02点)。对于所有CIS标本;通过肌腱缝合了,2标本失败缝合断裂。不同刚度之间的观察两组(CIS = 98.33 + / - 22.98 N / m vs集团=75.94 + / - 44.83 N / m;研究发现,打开我们的二头肌腱固定术执行集团建造结果在增加负载故障相比CIS。减少位移比集团这尸体生物力学的研究。相关性:集团和CIS提供时间为零固定的供应商术为LHBT腱固定术。然而,去除骨密质比CIS,因此会产生较小的立管压力肱骨近端。集团的完整性是确定的的完整性与周期性腱固定术装载。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号