首页> 外文期刊>Applied radiation and isotopes: including data, instrumentation and methods for use in agriculture, industry and medicine >A preliminary study on using the radiochromic film for 2D beam profile QC/QA at the THOR BNCT facility.
【24h】

A preliminary study on using the radiochromic film for 2D beam profile QC/QA at the THOR BNCT facility.

机译:在THOR BNCT设施中对2D光束轮廓QC / QA使用放射致变色膜的初步研究。

获取原文
获取原文并翻译 | 示例
           

摘要

The GAFCHROMIC((R)) EBT2 dosimetry film has been studied as a rapid QC/QA tool for 2D dose profile mapping in the BNCT beam at THOR. The pixel values of the EBT2 film image were converted to the 2D dose profile using a dose calibration curve obtained by 6-MV X-ray. The reproducibility of the 2D dose profile measured using the EBT2 film in the PMMA phantom was preliminarily found to be acceptable with uncertainties within about +/-2 to +/-3.5%. It is found that the EBT2 measured dose profile consisted of both gamma-ray components and neutron contributions. Therefore, the dose profile measured using the EBT2 film is significantly different from the neutron flux profile measured using the indirect neutron radiography method. Further study of the influence of neutrons to the response of the EBT2 film is indispensible for the absolute dose profile determination in a BNCT beam.
机译:已经研究了GAFCHROMIC EBT2剂量测定膜作为用于在THOR处的BNCT束中二维剂量分布图的快速QC / QA工具。使用通过6-MV X射线获得的剂量校准曲线,将EBT2胶片图像的像素值转换为2D剂量分布。初步发现使用EBT2膜在PMMA体模中测量的2D剂量分布的可重复性是可接受的,不确定性在+/- 2至+/- 3.5%范围内。发现EBT2测得的剂量分布由伽马射线成分和中子贡献组成。因此,使用EBT2膜测量的剂量分布与使用间接中子射线照相法测量的中子通量分布显着不同。对于BNCT束中绝对剂量分布的测定,中子对EBT2膜响应的影响的进一步研究是必不可少的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号