Evaluation of the Partec flow cytometer against the BD FACSCalibur system for monitoring immune responses of human immunodeficiency virus-infected patients in Zimbabwe.

机译:评估Partec流式细胞分析仪BD FACSCalibur免疫监测系统反应人类免疫缺陷感染病毒的病人在津巴布韦。

获取原文
获取原文并翻译 | 示例
           

摘要

A single-platform volumetric flow cytometer, the Partec Cyflow SL_3, was evaluated against a BD FACSCalibur/Sysmex XT1800i dual platform for measuring CD4(+) lymphocytes, total lymphocytes, and the percentage of CD4 lymphocytes in whole-blood samples for monitoring the immune systems of human immunodeficiency virus (HIV)/AIDS patients. Statistical analyses for precision, correlation, and agreement were performed. Coefficients of variation (CV) of 5.8, 4.6, and 3.9% were obtained for low, medium, and high CD4(+) cell counts, respectively, using the SL_3, and CV of 3.7, 4.0, and 0.94 were obtained for the same categories, using the BD FACSCalibur. Significant correlations (P < 0.005) between the two assays for CD4 counts, total lymphocyte counts, and percentages of CD4 were obtained, with correlation coefficients of 0.99, 0.96, and 0.99, respectively (n = 229). Using the Bland-Altman plot, mean biases of -18 cell/microl (95% confidence interval (CI); -91 to 54 cells/microl), -0.8% (95% CI; -3.6 to 2%), and -36.8 cells/microl (95% CI; -477 to 404 cells/microl) were obtained for comparisons of CD4 counts, percentages of CD4 cells, and total lymphocyte counts, respectively. The effects of the age of the samples on the three parameters were also analyzed by comparing results from the same samples analyzed at 6, 24, and 48 h after collection. The correlation coefficients for comparisons among different time points for the same machine and among all the time points for the two different machines were greater than 0.90. These data showed that the Partec Cyflow SL_3 assay is comparable to the BD FACSCalibur/Sysmex XT1800i dual-platform method for measuring the amount of CD4(+) cells and total lymphocytes and the percentages of CD4 cells in blood samples for the purpose of monitoring HIV/AIDS patients.
机译:一个单一平台体积流式细胞分析仪,对BD Partec Cyflow SL_3,评估FACSCalibur / Sysmex XT1800i双平台测量CD4(+)淋巴细胞,淋巴细胞,和CD4淋巴细胞的比例遗传免疫监测样本人类免疫缺陷病毒的系统艾滋病毒/艾滋病患者。精度、相关性和协议执行。4.6, 3.9%的人获得了低、中、和高CD4(+)细胞计数,分别使用SL_3简历为3.7,4.0和0.94为同一类别,使用双相障碍FACSCalibur。之间的两个化验CD4计数,总数的CD4淋巴细胞计数和百分比,相关系数为0.99,分别为0.96和0.99 (n = 229)。Bland-Altman情节,意味着-18细胞/ microl的偏见(95%可信区间(CI);细胞/ microl)、-0.8%(95%可信区间;-36.8细胞/ microl(95%可信区间;细胞/ microl)得到的比较CD4细胞的CD4计数,百分比,总数淋巴细胞计数,分别。样本的年龄三个参数也分析了比较的结果吗相同的样品分析在6、24和48 h后收集。比较在不同的时间点同一台机器,在所有时间点大于两个不同的机器上0.90. SL_3分析与双相障碍FACSCalibur / Sysmex XT1800i双平台的方法测量CD4(+)细胞的数量总淋巴细胞和CD4的百分比细胞在血液样本的目的监测艾滋病毒/艾滋病患者。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号