Biomechanical strain analysis of the proximal femur after retrograde intramedullary nailing

机译:生物力学应变分析的近端后股骨逆行髓内钉

获取原文
获取原文并翻译 | 示例
           

摘要

Background: The purpose of this study was to examine the influence of proximal retrograde intramedullary nail position on proximal femoral strain, since stress risers occurring at the end of an implant can increase fracture risk. Methods: Proximal femoral strains during axial and torsional loading were measured in composite Sawbone femurs after placement of retrograde intramedullary nails that ended at three different locations (2 cm proximal, 4 cm distal, and at the level of the lesser trochanter). Results: No statistically significant difference was found between the axial or torsional strain observed in the intact femur and that seen after placement of a retrograde femoral nail ending at any of the three positions. Gauges proximal to the nail tip demonstrated higher strains than the strains for the intact femur when compared with gauges distal to the nail tip. Conclusion: The ending location of a retrograde nail in the proximal femur does not appear to significantly alter strain in the proximal femur under the axial and torsional loading methods used in the study.
机译:背景:本研究的目的检查近端逆行的影响髓内钉在股骨近端上的立场压力,因为压力感应器发生的植入可以增加骨折的风险。方法:近端股骨在轴向压力在复合和扭转加载测量Sawbone股骨逆行位置后髓内钉在三结束近端不同位置(2厘米,4厘米远,和在小转子)。结果:没有统计上的显著差异被发现之间的轴向或扭转应变观察到完整的股骨,看过之后放置一个逆行股钉结束的三个位置。指甲尖了更高的压力比菌株的完整股骨相比仪表远端指甲尖。结束逆行钉的位置近端股骨似乎并不明显改变应变下的近端股骨轴向和扭转加载方法中使用研究。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号