Sources of out?of?field dose in MRgRT: an inter?comparison of measured and Monaco treatment planning system doses for the Elekta Unity MR?linac

机译:的来源? ?国际米兰?计划系统剂量Elekta团结先生吗?

获取原文
获取原文并翻译 | 示例
           

摘要

With the clinical introduction of MR-linacs, out-of-field dose (OFD) associated with head leakage/scatter (HLS), spiralling contaminant electrons (SCE) and the electron streaming effect (ESE) is of interest. To investigate HLS and SCE, EBT3 film on solid water 5.0 cm beyond each edge of a 10.0 x 10.0 cm(2) field was used to determine depth-dose for 0 T and 1.5 T, in the isocentric plane. Additionally, ESE induced by the anterior imaging coil was quantified and the experimental arrangements to measure SCE and ESE were modelled using Monaco. For a clinical treatment of supraclavicular nodal disease, Monaco OFD was compared to in vivo measurements. For 0 T, depth-dose was isotropic and surface dose was approximately 4.4% of D-max. With 1.5 T surface doses were approximately 3.8% of D-max at +/- Y (IEC61217), compared to 2.6% and 0.6% of D-max at - X and X, respectively. For both field strengths, the TPS depth-dose variation was consistent with experimental trends; however, near surface doses calculated at +/- Y differed significantly from measurements. For the field sizes investigated, measured coil ESE dose was between 9.0 and 28.0% of D-max and Monaco coil ESE was less than measured by up to 13.0%. OFD in 0 T and 1.5 T are comparable at +/- Y, inconsistent with previous work. Anterior coil ESE should be mitigated during treatment and for the clinical case investigated, in vivo OFD was within 2 sigma of TPS calculations. Monaco overestimates near surface SCE and underestimates coil ESE.
机译:与临床MR-linacs介绍,out-of-field剂量(海底钻井)与头相关联泄漏/散射(HLS)、螺旋上升的污染物电子(SCE)和电子流效果(ESE)是感兴趣的。EBT3电影固体水5.0厘米以外的每条边10.0 x 10.0厘米(2)字段被用来确定为0和1.5 T,深度剂量isocentric平面。前成像线圈和量化实验安排测量SCE和东南偏东模拟使用摩纳哥。锁骨上淋巴结疾病的治疗,摩纳哥海底钻井与体内测量。0 T,深度剂量是各向同性和表面剂量是D-max的大约4.4%。表面剂量D-max的大约3.8%+ / - Y (IEC61217),而2.6%和0.6%D-max - X, X,分别。优势,TPS深度剂量变化与实验一致的趋势;近地表在+ / - Y不同剂量计算明显的测量。尺寸调查,测量线圈ESE剂量28.0%和9.0之间D-max和摩纳哥线圈ESE不到13.0%的测量。0和1.5 T + / - Y可比,与以前的工作不一致。ESE应该在治疗和减轻临床病例研究,体内海底钻井在2σTPS的计算。高估了近地表SCE和低估coil大叫一声

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号